Filmetrics F20 Standard Operating Procedure
Jump to navigation
Jump to search
Measuring film thickness
- Open the Filmetrics software on the desktop.
- Perform a baseline:
- Focus the camera on the calibration wafer:
- Place the square area labeled "focus region" of the Filmetrics-branded 4" calibration wafer under the optical head of the tool.
- Click on the live video tab
- Use the silver vertical focus knob to focus the video until the pattern is sharp.
- Click "baseline"
- When prompted, adjust the calibration wafer such that the square, not-patterned silicon region in the center of the wafer is under the optical head.
- When prompted, remove the calibration wafer and insert the sample to be measured.
- If the sample is too small to sit flat on the center of the stage, place it on a carrier wafer.
- Re-adjust focus onto the top surface
- When prompted, insert the small 45-degree black reflector for the background measurement (this reflects light away from the return path)
- No focus change is necessary.
- Focus the camera on the calibration wafer:
- Measure your sample:
- Switch to the Measure tab
- In the right-hand pane, select the appropriate recipe that describes your sample.
- Click the Measure button.
- The software should report a thickness and a fit goodness.
- If the fit goodness is less than 97%, question either your recipe selection or the sample.
- Ctrl + M will enable continuous measurement, the same will disable it.