XeF2 Etcher SOP: Difference between revisions

From USC Nanofab Wiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 12: Line 12:
   separately (folder "XeF2_SOP_images"). Upload each one to the wiki via
   separately (folder "XeF2_SOP_images"). Upload each one to the wiki via
   Special:Upload using the EXACT file name shown below, so the [[File:...]] tags
   Special:Upload using the EXACT file name shown below, so the [[File:...]] tags
   resolve into the actual pictures instead of red "file missing" links.
   resolve into the actual pictures.


   Layout notes (this revision):
   Layout notes (this revision):
   - The Operating Procedure is now ONE continuous <ol> (steps 1-17) instead of
  - Images are now centered, borderless, uncaptioned blocks placed directly under
     several separate list fragments, so numbering is generated by the wiki
    the step/paragraph they follow -- the same way each image sits centered on its
    software itself rather than being stitched together manually.
    own beneath the text in the PDF (no floating/wrapping, no thumbnail frame or
  - Each photo is now placed inside the <li> of the exact step it illustrates in
    caption box, since the PDF itself has none).
    the PDF (as a right-floating thumbnail), so it sits next to that step instead
   - The Operating Procedure is one continuous <ol> (steps 1-17) so the numbering
    of appearing as a separate block between steps.
     is generated by the wiki software, not stitched together manually.
   - HTML comments mark where each PDF page began, so the page order still mirrors
   - HTML comments mark where each PDF page began, so the page order still mirrors
     the original 5-page document even though it's one continuous wiki page.
     the original 5-page document even though it's one continuous wiki page.
Line 32: Line 32:


== Standard Operating Procedure: XeF2 Etcher ==
== Standard Operating Procedure: XeF2 Etcher ==
[[File:XeF2_SOP_Cover_Photo.jpg|thumb|right|300px|Photo from the SOP cover page]]
 
[[File:XeF2_SOP_Cover_Photo.jpg|center|300px]]


''In case of emergency, call DPS: 213-740-4321''
''In case of emergency, call DPS: 213-740-4321''
Line 54: Line 55:
== MACHINE DESCRIPTION ==
== MACHINE DESCRIPTION ==
     Refer to Figure 1 for machine configuration and valving. There are 5 valves referred to as V1, V2, V3, V4, and V5. Their functions are:
     Refer to Figure 1 for machine configuration and valving. There are 5 valves referred to as V1, V2, V3, V4, and V5. Their functions are:
[[File:XeF2_SOP_Figure1_LabVIEW_Panel.png|thumb|right|400px|Figure 1 – Machine configuration on LabVIEW front panel]]


{| class="wikitable"
{| class="wikitable"
Line 69: Line 68:
| V5 || XeF2 source valve which allows vapor from the crystals to flow into the pulse chamber
| V5 || XeF2 source valve which allows vapor from the crystals to flow into the pulse chamber
|}
|}
[[File:XeF2_SOP_Figure1_LabVIEW_Panel.png|center|400px]]
<div style="text-align:center">Figure 1 – Machine configuration on LabVIEW front panel</div>


'''"Process Control" Section – Parameter Descriptions'''
'''"Process Control" Section – Parameter Descriptions'''
Line 108: Line 110:
== OPERATING PROCEDURE ==
== OPERATING PROCEDURE ==
<ol>
<ol>
<li>Check to see that XeF2 valve is closed (Note: the arrows point towards each other, “→ ←”)
<li>Check to see that XeF2 valve is closed (Note: the arrows point towards each other, “→ ←”)</li>
[[File:XeF2_SOP_Valve_Closed.jpg|thumb|right|250px|XeF2 valve closed]]</li>
</ol>
[[File:XeF2_SOP_Valve_Closed.jpg|center|250px]]
<ol start="2">
<li>Activate the tool on NEMO</li>
<li>Activate the tool on NEMO</li>
<li>Check to see that the tool is in ‘Standby’ mode
<li>Check to see that the tool is in ‘Standby’ mode</li>
[[File:XeF2_SOP_LabVIEW_Standby_Arrow.jpg|thumb|right|350px|LabVIEW front panel showing the tool in Standby mode]]</li>
</ol>
[[File:XeF2_SOP_LabVIEW_Standby_Arrow.jpg|center|400px]]
<ol start="4">
<li>Fill in relevant information to the log sheet</li>
<li>Fill in relevant information to the log sheet</li>
<li>Click ‘Remove Wafer’, unscrew black knob, and wait until step is complete
<li>Click ‘Remove Wafer’, unscrew black knob, and wait until step is complete</li>
[[File:XeF2_SOP_Wafer_Removal.jpg|thumb|right|250px|Removing the black knob to access the wafer]]</li>
</ol>
[[File:XeF2_SOP_Wafer_Removal.jpg|center|280px]]


<!-- PDF page 5 -->
<!-- PDF page 5 -->
<ol start="6">
<li>Open chamber, load sample(s), close chamber, then tighten black knob</li>
<li>Open chamber, load sample(s), close chamber, then tighten black knob</li>
<li>Click ‘Purge’ and wait until step is complete<br>Note: Moisture that originated from ambient air will enter the chamber when you load your sample, so it is very important to remove this moisture with a sufficiently long purge step.</li>
<li>Click ‘Purge’ and wait until step is complete<br>Note: Moisture that originated from ambient air will enter the chamber when you load your sample, so it is very important to remove this moisture with a sufficiently long purge step.</li>
<li>Open the XeF2 valve (Note: the arrows do not point towards each other, “↑ ←”)
<li>Open the XeF2 valve (Note: the arrows do not point towards each other, “↑ ←”)</li>
[[File:XeF2_SOP_Valve_Open.jpg|thumb|right|250px|XeF2 valve open]]</li>
</ol>
[[File:XeF2_SOP_Valve_Open.jpg|center|250px]]
<ol start="9">
<li>Click ‘Pulse Etch’ and wait until step is complete<br>Note: The first few pulses will have difficulty in precisely attaining the desired etching pressure, but later pulses will occur at desired pressure.</li>
<li>Click ‘Pulse Etch’ and wait until step is complete<br>Note: The first few pulses will have difficulty in precisely attaining the desired etching pressure, but later pulses will occur at desired pressure.</li>
<li>Close the XeF2 valve (Note: the arrows point towards each other, “→ ←”)
<li>Close the XeF2 valve (Note: the arrows point towards each other, “→ ←”)</li>
[[File:XeF2_SOP_Valve_Closed.jpg|thumb|right|250px|XeF2 valve closed]]</li>
</ol>
[[File:XeF2_SOP_Valve_Closed.jpg|center|250px]]
<ol start="11">
<li>Click ‘Remove Wafer’ and wait until step is complete</li>
<li>Click ‘Remove Wafer’ and wait until step is complete</li>
<li>Unscrew the black knob and remove your sample(s)</li>
<li>Unscrew the black knob and remove your sample(s)</li>

Revision as of 12:24, 17 September 2026

Standard Operating Procedure: XeF2 Etcher

In case of emergency, call DPS: 213-740-4321

Written by Eugene Yoon

June 2023

INTRODUCTION

   The XeF2 etcher in the USC Nanofab lab uses XeF2 crystals and a vacuum chamber to produce fluorine vapors which isotropically etches silicon with very high selectivity to photoresist, oxides, nitrides and many metals. Typical silicon etch rates are 1-2 µm per minute but can vary widely from loading effects. Samples up to 6” diameter can be accommodated. Typical applications are for cantilever release etching for micro-machined structures.

WARNINGS

   Although XeF2 itself is not considered very toxic, the production of HF vapors by exposing the XeF2 crystals to air is a very serious potential health threat. Only trained personnel may handle the crystals directly. Observe all purge and pumping procedures. If any odors are detected, put the system in standby and contact staff immediately. Do NOT proceed with etching. HF fumes have a pungent, irritating, penetrating odor and can be smelled at 0.04 ppm (a relatively safe level) though you should not rely on odor to determine safe concentrations since the odor threshold will vary widely with individuals
   The crystals are also very expensive and react with water vapor. The XeF2 source chamber should never be exposed to air and should be isolated when etching is complete to prevent the crystals from being pumped away.

SAMPLE PREPARATION

   Samples must be clean, dry, and free from oxide otherwise etching will be slow and non-uniform. An HF dip followed by a rinse and dehydration bake may be required to ensure adequate cleanliness and dryness. Oxides, photoresist, and nitrides provide high selectivity masking to this etch. Loading effects may be observed depending on sample size and amount of exposed silicon. Always perform a test run with the exact same sample size and pattern when determining etch rate and depth. In general, better results are obtained by etching in one run rather than several shorter runs where the sample is removed and then re-etched.

MACHINE DESCRIPTION

   Refer to Figure 1 for machine configuration and valving. There are 5 valves referred to as V1, V2, V3, V4, and V5. Their functions are:
V1 Main vacuum valve which allows pumping of the etching chamber. There should always be a small N2 flow during standby mode to prevent oil backstreaming.
V2 Flows nitrogen into the etch chamber for purging to remove XeF2 vapors and for back-filling the chamber so that the lid can be opened.
V3 Allows flow of XeF2 vapor from pulsing chamber into the etch chamber where the sample resides
V4 Flows nitrogen to the pulsing chamber
V5 XeF2 source valve which allows vapor from the crystals to flow into the pulse chamber
Figure 1 – Machine configuration on LabVIEW front panel

"Process Control" Section – Parameter Descriptions

Parameter Description
Etching Pressure Pressure value setpoint of etching chamber during pulse etches
Pumping Pressure Base pressure when evacuating the etching chamber
Duration Time Amount of time spent in each etching pulse
Purge Duration Amount of time spent in each purging cycle
Number of Pulses Number of times there will be etching pulses
N2 Purge Cycles Number of times there will be N2 purge cycles

BACKGROUND + ADDITIONAL INFO

   At room temperature, XeF2 is in a solid crystalline form. When it is put onto low pressure, XeF2 will sublimate into gas. This gaseous XeF2 can etch silicon. It does not etch silica, most photoresists, oxides, nitrides, or many metals. A more detailed explanation of the etching mechanism can be found in B. Garrison and W. Goddard’s “Reaction Mechanism for Fluorine Etching of Silicon” Physical Review B 36:18 (1987)
   The Etching Chamber is located at the top of the etcher and can be opened and closed to load samples using the black knob. It is possible to operate the etcher manually with the 5 vacuum switches and pressure display, but users must operate the tooling using automatic mode. This ensures accurate operation and saves time.
   The Pressure Control Valve controls the rate at which the etching chamber’s pressure drops. To prevent lightweight samples from flipping over or blowing away, turn the pressure control valve to lower the rate at which the pressure drops.
   If you make an error, press the red “Stop” button in the top left corner of the LabVIEW program. Then, press the “Run” button in the top left corner of the LabVIEW program. Click “Purge” to purge the chamber.
   To restart the LabVIEW program, press the “Stop” button in the top left corner of the LabVIEW program. Close the program altogether by clicking the red and white “X” on the top right corner. Open the file called “USCxef210-8-07source” on the desktop. Then, press the “Run” button in the top left corner of the LabVIEW program. Then click “Standby” to bring the tool back to Standby mode.

Please notify staff if:

  1. The XeF2 valve was left open
  2. The system’s initial etching chamber pressure is over 100 mTorr (it should be ~20 mTorr)
  3. The system takes a long time to reach its etching pressure during pulse etch

OPERATING PROCEDURE

  1. Check to see that XeF2 valve is closed (Note: the arrows point towards each other, “→ ←”)
  1. Activate the tool on NEMO
  2. Check to see that the tool is in ‘Standby’ mode
  1. Fill in relevant information to the log sheet
  2. Click ‘Remove Wafer’, unscrew black knob, and wait until step is complete
  1. Open chamber, load sample(s), close chamber, then tighten black knob
  2. Click ‘Purge’ and wait until step is complete
    Note: Moisture that originated from ambient air will enter the chamber when you load your sample, so it is very important to remove this moisture with a sufficiently long purge step.
  3. Open the XeF2 valve (Note: the arrows do not point towards each other, “↑ ←”)
  1. Click ‘Pulse Etch’ and wait until step is complete
    Note: The first few pulses will have difficulty in precisely attaining the desired etching pressure, but later pulses will occur at desired pressure.
  2. Close the XeF2 valve (Note: the arrows point towards each other, “→ ←”)
  1. Click ‘Remove Wafer’ and wait until step is complete
  2. Unscrew the black knob and remove your sample(s)
  3. If you have more samples, load them and repeat the procedure from step 6
  4. If you are finished etching, close the chamber and tighten the black knob, and click ‘Standby’
  5. Wait 2 minutes
  6. Fill in the ‘End Status’ on the log sheet and add any comments in ‘Notes’ (if you have any)
  7. Deactivate tool on NEMO

In case of emergency, call DPS: 213-740-4321

Written by Eugene Yoon

June 2023