Ellipsometer Standard Operating Procedure: Difference between revisions
Jump to navigation
Jump to search
Add screenshot |
Add photo |
||
| Line 1: | Line 1: | ||
[[File:Ellipsometer components.png|thumb|500x500px|Ellipsometer components needed for operation.]] | |||
== Ellipsometer Standard Operating Procedure == | == Ellipsometer Standard Operating Procedure == | ||
| Line 12: | Line 14: | ||
# Z adjust procedure: | # Z adjust procedure: | ||
## Click "Adjust Sample Table" | ## Click "Adjust Sample Table" | ||
## Unlock the Z | ## Unlock the Z lock screw. | ||
##* It is silver, knurled, and pointing out at the user from under the sample table. | ##* It is silver, knurled, and pointing out at the user from under the sample table. | ||
## Pull out the beam stop to allow the red laser beam to strike the sample table. | ## Pull out the beam stop to allow the red laser beam to strike the sample table. | ||
## Use the 3" knurled ring under the sample table to adjust the table height | ## Use the 3" knurled ring under the sample table to adjust the table height | ||
##* Do this while monitoring the Table Height Adjustment view in the software until the table position is centered. | ##* Do this while monitoring the Table Height Adjustment view in the software until the table position is centered. | ||
## Lock the Z | ## Lock the Z lock screw and ensure the table alignment has not shifted. | ||
## Click OK to exit the table adjustment dialog. | ## Click OK to exit the table adjustment dialog. | ||
# Check the ellipsometer function with a calibration standard for your film(s) if available: | # Check the ellipsometer function with a calibration standard for your film(s) if available: | ||
Revision as of 15:18, 28 March 2025

Ellipsometer Standard Operating Procedure
- Turn the keyswitch on the back-right clockwise to turn the tool on.
- On the ellipsometer computer, open up GEMP program, then go to the menu option Ellipsometer > Waferskan Mapping
- Align the table tilt and height:
- Place any piece of flat reflective substrate on the table.
- Ensure that the laser beam stop is pushed in to protect your eyes.
- It is silver, knurled, and pointing out at the user from the body of the laser to the left of and above the sample table.
- Look through microscope and us vertical XY knobs on the black round plate to center the two crosshairs with the microscope.
- Place either a calibration standard, or the sample to be measured on the center of the ellipsometer table.
- Z adjust procedure:
- Click "Adjust Sample Table"
- Unlock the Z lock screw.
- It is silver, knurled, and pointing out at the user from under the sample table.
- Pull out the beam stop to allow the red laser beam to strike the sample table.
- Use the 3" knurled ring under the sample table to adjust the table height
- Do this while monitoring the Table Height Adjustment view in the software until the table position is centered.
- Lock the Z lock screw and ensure the table alignment has not shifted.
- Click OK to exit the table adjustment dialog.
- Check the ellipsometer function with a calibration standard for your film(s) if available:
- In this example, we will use the SiO2 1990 Angstrom standard from Gaertner.
- Return to the Waferskan Mapping window.
- Load a model:
- Click "Load New Model File" in the "Thin Film Model" pane.
- Select a TFM file and click "Open."
- i.e., "Thin Oxide.tfm" for SiO2.
- In the field "Thickness 1" in the "Measurement and Calculation" pane, enter the expected thickness of the sample (in Angstroms)
- i.e., 1900
- Know which wavelength the tool is using in the "Thin Film Model" frame of the Waferscan Mapping window.
- This will either be 4050, 6328, or 8300 Angstroms.
- This wavelength can be changed between the available options by clicking on the "Wavelength" button
- The wavelengths can also be disabled by unchecking the box for that wavelength.
- Click "Meas & Calc" in the "Measurement and Calculation" pane.
- Note that there are separate buttons for these steps as well:
- The "Measure" step:
- This will get new values for the Psi and Delta fields.
- i.e., Psi=33.341, Delta=279.711
- The "Calculate" step:
- This will display a thickness under the "Thick1" field, and an index of refraction under the "Nf1" field, as well as any other layers present in the model.
- i.e., Thick1=1976.951, Nf1=1.4706
- The "Measure" step:
- Note that there are separate buttons for these steps as well:
- If the measurement is OK and matches the expected standard thickness, proceed to measure your samples.
- Switch over to your sample on the sample table and measure:
- Ensure that the laser beam stop is pushed in to protect your eyes.
- Adjust table tilt and table height again.
- Enter expected thickness of the film
- Enter expected index of refraction of the film
- Change to the desired measurement wavelength.
- Click "Meas & Calc" as before to acquire the calculated thickness.
- Wrap up:
- Close software
- Turn off keyswitch on the back-right
Creating thin film models for the ellipsometer
- On the ellipsometer computer, open up GEMP program, then go to the menu option Ellipsometer > Waferskan Mapping
- Look up the indices of refraction for your sample substrate and films for the wavelengths 4050, 6328, and 8300 Angstroms.
- These may be found in a table in the green-bound ellipsometer manual.
- These may be measured on the F20 tool.
- https://refractiveindex.info/ is also good resource.
- Click "Save New Model File" in the "Thin Film Model" pane.
- Save the name of your new model in the model directory, descriptively titled in the format "[film material] on [substrate material].tfm"
- You can also specify the source for your constants in the filename.
- Check which wavelength the tool is using in the "Thin Film Model" frame of the Waferscan Mapping window.
- This will either be 4050, 6328, or 8300 Angstroms.
- This wavelength can be changed between the available options by clicking on the "Wavelength" button
- The wavelengths can also be disabled by unchecking the box for that wavelength.
- Enter data into the model file:
- Enter estimated thicknesses under the "Thick" field. index of refraction active wavelength under the "Nf" and "Kf" fields
- For the thin films, enter data in the numbered rows, starting at the top row in the "Thin Film Model" pane, up to 4 films possible.
- For the substrate, enter data at the bottom row labeled "S"
- Cycle through each of the three wavelengths, repeating data entry for each.
- Save again the name of your new model in the model directory, overwriting the same file you created earlier.
- The model can now be used for measurement and recalled when needed.