Created page with "<!-- New wiki page: SOP content transcribed from "XeF2 SOP - June 2023.pdf" Original SOP written by: Eugene Yoon, June 2023 Transcribed to wiki by: Chandan Ramakrishnaiah Role: Lab Operation Engineer (Staff) Phone: +1 (213) 551-6726 Email: chandanr@usc.edu SUGGESTED PAGE TITLE: "XeF2 Etcher SOP" (rename this title to match your wiki's naming convention if different, and update the link on the main XeF2 equipment page to match) --> __TOC__ == Overview =..."
 
No edit summary
Line 1: Line 1:
<!--
<!--
   New wiki page: SOP content transcribed from "XeF2 SOP - June 2023.pdf"
   New wiki page: exact transcription of "XeF2 SOP - June 2023.pdf" (text and images),
   Original SOP written by: Eugene Yoon, June 2023
   written by Eugene Yoon, June 2023.
   Transcribed to wiki by: Chandan Ramakrishnaiah
   Transcribed by: Chandan Ramakrishnaiah
   Role: Lab Operation Engineer (Staff)
   Role: Lab Operation Engineer (Staff)
   Phone: +1 (213) 551-6726
   Phone: +1 (213) 551-6726
Line 8: Line 8:


   SUGGESTED PAGE TITLE: "XeF2 Etcher SOP"
   SUGGESTED PAGE TITLE: "XeF2 Etcher SOP"
   (rename this title to match your wiki's naming convention if different,
 
   and update the link on the main XeF2 equipment page to match)
   IMAGES: the 7 image files referenced below (File:XeF2_SOP_...) are included as
  separate files alongside this one (folder "XeF2_SOP_images"). Upload each one to
  the wiki with Special:Upload using the exact file name shown, so the [[File:...]]
  links resolve. They are the unedited images extracted directly from the PDF.
 
  Note on the PDF's running header/footer: the USC logo banner and the line
  "In case of emergency, call DPS: 213-740-4321 / Written by Eugene Yoon / June 2023"
  appeared identically at the top/bottom of all 5 PDF pages (it is a repeating
   page header/footer, not page content). It is placed once at the top and once at
  the bottom here since this is a single continuous wiki page rather than 5 printed
  pages; no wording was changed.
-->
-->
__TOC__
__TOC__
== Overview ==
[[File:XeF2_SOP_USC_Logo_Banner.png|200px]]
'''Standard Operating Procedure: XeF₂ Etcher'''


The XeF₂ etcher in the USC Nanofab lab uses XeF₂ crystals and a vacuum chamber to produce fluorine vapors which isotropically etch silicon with very high selectivity to photoresist, oxides, nitrides, and many metals. Typical silicon etch rates are 1–2 µm per minute but can vary widely due to loading effects. Samples up to 6" diameter can be accommodated. Typical applications are cantilever release etching for micro-machined structures.
== Standard Operating Procedure: XeF2 Etcher ==
[[File:XeF2_SOP_Cover_Photo.jpg|thumb|300px|Photo from the SOP cover page]]


''Originally written by Eugene Yoon, June 2023. See also the main equipment page: [[XeF2|XeF₂ Etcher]].''
''In case of emergency, call DPS: 213-740-4321''


== Warnings ==
''Written by Eugene Yoon''
{| class="wikitable" style="width:100%; margin-top:12px; border:3px solid #990000; background:#ffffff;"
 
|-
''June 2023''
! style="background:#990000; color:#FFCC00; padding:10px; font-size:14px;" | ⚠ Safety Warnings
|-
| style="padding:12px; background:#fff5f5;" |
Although XeF₂ itself is not considered very toxic, the production of '''hydrofluoric acid (HF) vapors''' by exposing XeF₂ crystals to air is a serious potential health threat.


* Only trained personnel may handle the crystals directly.
== INTRODUCTION ==
* Observe all purge and pumping procedures.
    The XeF2 etcher in the USC Nanofab lab uses XeF2 crystals and a vacuum chamber to produce fluorine vapors which isotropically etches silicon with very high selectivity to photoresist, oxides, nitrides and many metals. Typical silicon etch rates are 1-2 µm per minute but can vary widely from loading effects. Samples up to 6” diameter can be accommodated. Typical applications are for cantilever release etching for micro-machined structures.
* '''If any odors are detected, put the system in standby and contact staff immediately. Do NOT proceed with etching.'''
* HF fumes have a pungent, irritating, penetrating odor and can be smelled at 0.04 ppm (a relatively safe level) — but odor should not be relied on to judge safe concentration, since the odor threshold varies widely between individuals.
* The crystals are also very expensive and react with water vapor. The XeF₂ source chamber should never be exposed to air, and should be isolated when etching is complete to prevent the crystals from being pumped away.
|}


'''In case of emergency, call USC DPS: 213-740-4321'''
== WARNINGS ==
    Although XeF2 itself is not considered very toxic, the production of HF vapors by exposing the XeF2 crystals to air is a very serious potential health threat. Only trained personnel may handle the crystals directly. Observe all purge and pumping procedures. If any odors are detected, put the system in standby and contact staff immediately. Do NOT proceed with etching. HF fumes have a pungent, irritating, penetrating odor and can be smelled at 0.04 ppm (a relatively safe level) though you should not rely on odor to determine safe concentrations since the odor threshold will vary widely with individuals


== Sample Preparation ==
    The crystals are also very expensive and react with water vapor. The XeF2 source chamber should never be exposed to air and should be isolated when etching is complete to prevent the crystals from being pumped away.
Samples must be clean, dry, and free from oxide, otherwise etching will be slow and non-uniform. An HF dip followed by a rinse and dehydration bake may be required to ensure adequate cleanliness and dryness.


* Oxides, photoresist, and nitrides provide high-selectivity masking for this etch.
== SAMPLE PREPARATION ==
* Loading effects may be observed depending on sample size and the amount of exposed silicon.
    Samples must be clean, dry, and free from oxide otherwise etching will be slow and non-uniform. An HF dip followed by a rinse and dehydration bake may be required to ensure adequate cleanliness and dryness. Oxides, photoresist, and nitrides provide high selectivity masking to this etch. Loading effects may be observed depending on sample size and amount of exposed silicon. Always perform a test run with the exact same sample size and pattern when determining etch rate and depth. In general, better results are obtained by etching in one run rather than several shorter runs where the sample is removed and then re-etched.
* Always perform a test run with the exact same sample size and pattern when determining etch rate and depth.
* In general, better results are obtained by etching in one run rather than several shorter runs where the sample is removed and re-etched.


== Machine Description ==
== MACHINE DESCRIPTION ==
Refer to the LabVIEW front panel (Figure 1) for machine configuration and valving. There are 5 valves, referred to as V1–V5:
    Refer to Figure 1 for machine configuration and valving. There are 5 valves referred to as V1, V2, V3, V4, and V5. Their functions are:


{| class="wikitable" style="width:100%;"
{| class="wikitable"
! Valve !! Function
|-
|-
| V1 || Main vacuum valve, allows pumping of the etching chamber. There should always be a small N₂ flow during standby mode to prevent oil backstreaming.
| V1 || Main vacuum valve which allows pumping of the etching chamber. There should always be a small N2 flow during standby mode to prevent oil backstreaming.
|-
|-
| V2 || Flows nitrogen into the etch chamber for purging (to remove XeF₂ vapors) and for back-filling the chamber so the lid can be opened.
| V2 || Flows nitrogen into the etch chamber for purging to remove XeF2 vapors and for back-filling the chamber so that the lid can be opened.
|-
|-
| V3 || Allows flow of XeF₂ vapor from the pulsing chamber into the etch chamber where the sample resides.
| V3 || Allows flow of XeF2 vapor from pulsing chamber into the etch chamber where the sample resides
|-
|-
| V4 || Flows nitrogen to the pulsing chamber.
| V4 || Flows nitrogen to the pulsing chamber
|-
|-
| V5 || XeF₂ source valve — allows vapor from the crystals to flow into the pulse chamber.
| V5 || XeF2 source valve which allows vapor from the crystals to flow into the pulse chamber
|}
|}


=== "Process Control" section – parameter descriptions ===
[[File:XeF2_SOP_Figure1_LabVIEW_Panel.png|thumb|400px|Figure 1 – Machine configuration on LabVIEW front panel]]
{| class="wikitable" style="width:100%;"
 
'''"Process Control" Section – Parameter Descriptions'''
 
{| class="wikitable"
! Parameter !! Description
! Parameter !! Description
|-
|-
| Etching Pressure || Pressure value setpoint of the etching chamber during pulse etches.
| Etching Pressure || Pressure value setpoint of etching chamber during pulse etches
|-
|-
| Pumping Pressure || Base pressure when evacuating the etching chamber.
| Pumping Pressure || Base pressure when evacuating the etching chamber
|-
|-
| Duration Time || Amount of time spent in each etching pulse.
| Duration Time || Amount of time spent in each etching pulse
|-
|-
| Purge Duration || Amount of time spent in each purging cycle.
| Purge Duration || Amount of time spent in each purging cycle
|-
|-
| Number of Pulses || Number of times there will be etching pulses.
| Number of Pulses || Number of times there will be etching pulses
|-
|-
| N₂ Purge Cycles || Number of times there will be N₂ purge cycles.
| N2 Purge Cycles || Number of times there will be N2 purge cycles
|}
|}


The etcher is operated through a LabVIEW front panel (program file: <code>USCxef210-8-07source</code>) showing the valve control indicators and process control parameters above.
== BACKGROUND + ADDITIONAL INFO ==
    At room temperature, XeF2 is in a solid crystalline form. When it is put onto low pressure, XeF2 will sublimate into gas. This gaseous XeF2 can etch silicon. It does not etch silica, most photoresists, oxides, nitrides, or many metals. A more detailed explanation of the etching mechanism can be found in B. Garrison and W. Goddard’s “Reaction Mechanism for Fluorine Etching of Silicon” Physical Review B 36:18 (1987)
 
    The Etching Chamber is located at the top of the etcher and can be opened and closed to load samples using the black knob. It is possible to operate the etcher manually with the 5 vacuum switches and pressure display, but users must operate the tooling using automatic mode. This ensures accurate operation and saves time.
 
    The Pressure Control Valve controls the rate at which the etching chamber’s pressure drops. To prevent lightweight samples from flipping over or blowing away, turn the pressure control valve to lower the rate at which the pressure drops.
 
    If you make an error, press the red “Stop” button in the top left corner of the LabVIEW program. Then, press the “Run” button in the top left corner of the LabVIEW program. Click “Purge” to purge the chamber.
 
    To restart the LabVIEW program, press the “Stop” button in the top left corner of the LabVIEW program. Close the program altogether by clicking the red and white “X” on the top right corner. Open the file called “USCxef210-8-07source” on the desktop. Then, press the “Run” button in the top left corner of the LabVIEW program. Then click “Standby” to bring the tool back to Standby mode.
 
Please notify staff if:
# The XeF2 valve was left open
# The system’s initial etching chamber pressure is over 100 mTorr (it should be ~20 mTorr)
# The system takes a long time to reach its etching pressure during pulse etch
 
== OPERATING PROCEDURE ==
# Check to see that XeF2 valve is closed (Note: the arrows point towards each other, “→ ←”)
 
[[File:XeF2_SOP_Valve_Closed.jpg|thumb|250px|XeF2 valve closed]]
 
<ol start="2">
<li>Activate the tool on NEMO</li>
<li>Check to see that the tool is in ‘Standby’ mode</li>
</ol>
 
[[File:XeF2_SOP_LabVIEW_Standby_Arrow.jpg|thumb|350px|LabVIEW front panel showing the tool in Standby mode]]
 
<ol start="4">
<li>Fill in relevant information to the log sheet</li>
<li>Click ‘Remove Wafer’, unscrew black knob, and wait until step is complete</li>
</ol>
 
[[File:XeF2_SOP_Wafer_Removal.jpg|thumb|250px|Removing the black knob to access the wafer]]


== Background + Additional Info ==
<ol start="6">
At room temperature, XeF₂ is a solid crystalline form. Under low pressure, XeF₂ sublimates into gas. This gaseous XeF₂ etches silicon; it does not etch silica, most photoresists, oxides, nitrides, or many metals. A more detailed explanation of the etching mechanism can be found in B. Garrison and W. Goddard, "Reaction Mechanism for Fluorine Etching of Silicon," ''Physical Review B'' 36:18 (1987).
<li>Open chamber, load sample(s), close chamber, then tighten black knob</li>
<li>Click ‘Purge’ and wait until step is complete<br>Note: Moisture that originated from ambient air will enter the chamber when you load your sample, so it is very important to remove this moisture with a sufficiently long purge step.</li>
<li>Open the XeF2 valve (Note: the arrows do not point towards each other, “↑ ←”)</li>
</ol>


The '''Etching Chamber''' is located at the top of the etcher and can be opened and closed to load samples using the black knob. Manual operation is possible using the 5 vacuum switches and pressure display, but users '''must''' operate the tool in automatic mode, which ensures accurate operation and saves time.
[[File:XeF2_SOP_Valve_Open.jpg|thumb|250px|XeF2 valve open]]


The '''Pressure Control Valve''' controls the rate at which the etching chamber's pressure drops. ''To prevent lightweight samples from flipping over or blowing away, turn the pressure control valve to lower the rate at which the pressure drops.''
<ol start="9">
<li>Click ‘Pulse Etch’ and wait until step is complete<br>Note: The first few pulses will have difficulty in precisely attaining the desired etching pressure, but later pulses will occur at desired pressure.</li>
<li>Close the XeF2 valve (Note: the arrows point towards each other, “→ ←”)</li>
</ol>


'''If you make an error:''' press the red "Stop" button in the top-left corner of the LabVIEW program, then press "Run." Click "Purge" to purge the chamber.
[[File:XeF2_SOP_Valve_Closed.jpg|thumb|250px|XeF2 valve closed]]


'''To restart the LabVIEW program:''' press "Stop," then close the program entirely by clicking the red-and-white "X" in the top-right corner. Open the file called <code>USCxef210-8-07source</code> on the desktop, press "Run," then click "Standby" to bring the tool back to Standby mode.
<ol start="11">
<li>Click ‘Remove Wafer’ and wait until step is complete</li>
<li>Unscrew the black knob and remove your sample(s)</li>
<li>If you have more samples, load them and repeat the procedure from step 6</li>
<li>If you are finished etching, close the chamber and tighten the black knob, and click ‘Standby’</li>
<li>Wait 2 minutes</li>
<li>Fill in the ‘End Status’ on the log sheet and add any comments in ‘Notes’ (if you have any)</li>
<li>Deactivate tool on NEMO</li>
</ol>


'''Please notify staff if:'''
----
# The XeF₂ valve was left open.
''In case of emergency, call DPS: 213-740-4321''
# The system's initial etching chamber pressure is over 100 mTorr (it should be ~20 mTorr).
# The system takes a long time to reach its etching pressure during pulse etch.


== Operating Procedure ==
''Written by Eugene Yoon''
# Check that the XeF₂ valve is closed (the arrows point towards each other: "→ ←").
# Activate the tool on NEMO.
# Check that the tool is in 'Standby' mode.
# Fill in relevant information on the log sheet.
# Click 'Remove Wafer', unscrew the black knob, and wait until the step is complete.
# Open the chamber, load sample(s), close the chamber, then tighten the black knob.
# Click 'Purge' and wait until the step is complete. ''Note: moisture from ambient air enters the chamber when loading a sample — it is important to remove this moisture with a sufficiently long purge step.''
# Open the XeF₂ valve (the arrows do '''not''' point towards each other: "↑ ←").
# Click 'Pulse Etch' and wait until the step is complete. ''Note: the first few pulses will have difficulty precisely attaining the desired etching pressure, but later pulses will occur at the desired pressure.''
# Close the XeF₂ valve (arrows point towards each other: "→ ←").
# Click 'Remove Wafer' and wait until the step is complete.
# Unscrew the black knob and remove your sample(s).
# If you have more samples, load them and repeat the procedure from step 6.
# If you are finished etching, close the chamber, tighten the black knob, and click 'Standby'.
# Wait 2 minutes.
# Fill in the 'End Status' on the log sheet and add any comments in 'Notes' (if applicable).
# Deactivate the tool on NEMO.


== Related Pages ==
''June 2023''
* [[XeF2|XeF₂ Etcher — equipment page]]
* [[:File:XeF2 Quick Start Guide - May 2023.pdf|XeF₂ Quickstart Guide]]


[[Category:SOPs]]
[[Category:SOPs]]
[[Category:Deposition Bay 1]]
[[Category:Deposition Bay 1]]

Revision as of 12:18, 17 September 2026

Standard Operating Procedure: XeF2 Etcher

Photo from the SOP cover page

In case of emergency, call DPS: 213-740-4321

Written by Eugene Yoon

June 2023

INTRODUCTION

   The XeF2 etcher in the USC Nanofab lab uses XeF2 crystals and a vacuum chamber to produce fluorine vapors which isotropically etches silicon with very high selectivity to photoresist, oxides, nitrides and many metals. Typical silicon etch rates are 1-2 µm per minute but can vary widely from loading effects. Samples up to 6” diameter can be accommodated. Typical applications are for cantilever release etching for micro-machined structures.

WARNINGS

   Although XeF2 itself is not considered very toxic, the production of HF vapors by exposing the XeF2 crystals to air is a very serious potential health threat. Only trained personnel may handle the crystals directly. Observe all purge and pumping procedures. If any odors are detected, put the system in standby and contact staff immediately. Do NOT proceed with etching. HF fumes have a pungent, irritating, penetrating odor and can be smelled at 0.04 ppm (a relatively safe level) though you should not rely on odor to determine safe concentrations since the odor threshold will vary widely with individuals
   The crystals are also very expensive and react with water vapor. The XeF2 source chamber should never be exposed to air and should be isolated when etching is complete to prevent the crystals from being pumped away.

SAMPLE PREPARATION

   Samples must be clean, dry, and free from oxide otherwise etching will be slow and non-uniform. An HF dip followed by a rinse and dehydration bake may be required to ensure adequate cleanliness and dryness. Oxides, photoresist, and nitrides provide high selectivity masking to this etch. Loading effects may be observed depending on sample size and amount of exposed silicon. Always perform a test run with the exact same sample size and pattern when determining etch rate and depth. In general, better results are obtained by etching in one run rather than several shorter runs where the sample is removed and then re-etched.

MACHINE DESCRIPTION

   Refer to Figure 1 for machine configuration and valving. There are 5 valves referred to as V1, V2, V3, V4, and V5. Their functions are:
V1 Main vacuum valve which allows pumping of the etching chamber. There should always be a small N2 flow during standby mode to prevent oil backstreaming.
V2 Flows nitrogen into the etch chamber for purging to remove XeF2 vapors and for back-filling the chamber so that the lid can be opened.
V3 Allows flow of XeF2 vapor from pulsing chamber into the etch chamber where the sample resides
V4 Flows nitrogen to the pulsing chamber
V5 XeF2 source valve which allows vapor from the crystals to flow into the pulse chamber
Figure 1 – Machine configuration on LabVIEW front panel

"Process Control" Section – Parameter Descriptions

Parameter Description
Etching Pressure Pressure value setpoint of etching chamber during pulse etches
Pumping Pressure Base pressure when evacuating the etching chamber
Duration Time Amount of time spent in each etching pulse
Purge Duration Amount of time spent in each purging cycle
Number of Pulses Number of times there will be etching pulses
N2 Purge Cycles Number of times there will be N2 purge cycles

BACKGROUND + ADDITIONAL INFO

   At room temperature, XeF2 is in a solid crystalline form. When it is put onto low pressure, XeF2 will sublimate into gas. This gaseous XeF2 can etch silicon. It does not etch silica, most photoresists, oxides, nitrides, or many metals. A more detailed explanation of the etching mechanism can be found in B. Garrison and W. Goddard’s “Reaction Mechanism for Fluorine Etching of Silicon” Physical Review B 36:18 (1987)
   The Etching Chamber is located at the top of the etcher and can be opened and closed to load samples using the black knob. It is possible to operate the etcher manually with the 5 vacuum switches and pressure display, but users must operate the tooling using automatic mode. This ensures accurate operation and saves time.
   The Pressure Control Valve controls the rate at which the etching chamber’s pressure drops. To prevent lightweight samples from flipping over or blowing away, turn the pressure control valve to lower the rate at which the pressure drops.
   If you make an error, press the red “Stop” button in the top left corner of the LabVIEW program. Then, press the “Run” button in the top left corner of the LabVIEW program. Click “Purge” to purge the chamber.
   To restart the LabVIEW program, press the “Stop” button in the top left corner of the LabVIEW program. Close the program altogether by clicking the red and white “X” on the top right corner. Open the file called “USCxef210-8-07source” on the desktop. Then, press the “Run” button in the top left corner of the LabVIEW program. Then click “Standby” to bring the tool back to Standby mode.

Please notify staff if:

  1. The XeF2 valve was left open
  2. The system’s initial etching chamber pressure is over 100 mTorr (it should be ~20 mTorr)
  3. The system takes a long time to reach its etching pressure during pulse etch

OPERATING PROCEDURE

  1. Check to see that XeF2 valve is closed (Note: the arrows point towards each other, “→ ←”)
XeF2 valve closed
  1. Activate the tool on NEMO
  2. Check to see that the tool is in ‘Standby’ mode
LabVIEW front panel showing the tool in Standby mode
  1. Fill in relevant information to the log sheet
  2. Click ‘Remove Wafer’, unscrew black knob, and wait until step is complete
Removing the black knob to access the wafer
  1. Open chamber, load sample(s), close chamber, then tighten black knob
  2. Click ‘Purge’ and wait until step is complete
    Note: Moisture that originated from ambient air will enter the chamber when you load your sample, so it is very important to remove this moisture with a sufficiently long purge step.
  3. Open the XeF2 valve (Note: the arrows do not point towards each other, “↑ ←”)
XeF2 valve open
  1. Click ‘Pulse Etch’ and wait until step is complete
    Note: The first few pulses will have difficulty in precisely attaining the desired etching pressure, but later pulses will occur at desired pressure.
  2. Close the XeF2 valve (Note: the arrows point towards each other, “→ ←”)
XeF2 valve closed
  1. Click ‘Remove Wafer’ and wait until step is complete
  2. Unscrew the black knob and remove your sample(s)
  3. If you have more samples, load them and repeat the procedure from step 6
  4. If you are finished etching, close the chamber and tighten the black knob, and click ‘Standby’
  5. Wait 2 minutes
  6. Fill in the ‘End Status’ on the log sheet and add any comments in ‘Notes’ (if you have any)
  7. Deactivate tool on NEMO

In case of emergency, call DPS: 213-740-4321

Written by Eugene Yoon

June 2023