Filmetrics F10 Standard Operating Procedure

From USC Nanofab Wiki
Revision as of 16:13, 23 May 2025 by J (talk | contribs) (Create page, make some edits)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

Please note

This SOP is a rough draft and not to be executed.

Measuring film thickness

  1. Open the Filmetrics software on the desktop.
  2. Perform a baseline:
    1. Focus the camera:
      1. Place the square area labeled "focus region" of the Filmetrics-branded 4" calibration wafer under the optical head of the tool.
      2. Click on the live video tab
      3. Use the silver vertical focus knob to focus the video until the pattern is sharp.
    2. Click "baseline"
      1. When prompted, adjust the calibration wafer such that the square silicon region is under the optical head.

       For any measurement silicon based, can keep that baseline for all measurements

       If you don't get a "saturation error," then no baseline is needed

       Re-baseline every hour - software will remind you

   Move sample to the SiO2 on Si green circle