Electroglas 4080X
About



The Electroglas 4080X is an automated wafer prober used for wafer-level electrical testing. It accurately places a wafer under a probe card, uses vision for alignment, and steps across die sites so a tester can measure devices and record/sort results.
Typical systems are modular (handler + prober control + vision), often configured for 150 mm or 200 mm wafers, with common options such as a hot chuck (temperature probing) and GPIB/serial I/O for integration with parametric/test equipment.
DCM-2 (display/control), MHM (material handler), PCM (prober control), PRM-3
Configured for 200 mm (8") wafers with a hot chuck option
Key Features:
- Wafer size: 100 mm (4″) – 200 mm (8″) standard; ring-carrier adapters up to 300 mm available
- Chuck type & temp: Gold-plated vacuum chuck, ambient → +130 °C with TC-2000 controller (standard “hot chuck”)
- Indexing accuracy/repeatability: ±4 µm X-Y, Z-resolution 0.25 mil (6.3 µm)
- Throughput: ≤ 1.5 s die-to-die at 5 mm step; lot-level throughput ≈ 60 wph (150 mm)
- Vision/alignment: PRM-3 CCD camera, Self-Teach Auto Alignment (STAA), optional OCR & Probe-Mark-Inspection (PMI)
- Interfaces: GPIB-IEEE 488, RS-232, Ethernet, EG-Commander/ProberBench software
Applications:
- Wafer loading, precise alignment, and stepping for parametric/TEG testing
- Integration with external testers (e.g., Keithley S530) for automated electrical characterization
- Temperature-controlled probing for device performance at elevated temps
- High-accuracy die-to-die positioning in production monitoring flows
Detailed Specifications
Model: Electroglas 4080X Automated Wafer Prober Configured for 200 mm (8") wafers with a hot chuck option Location: [Specify lab bay if known, e.g., Test Bay]
Capability → Typical value/option
- Wafer size → 100 mm (4″) – 200 mm (8″) standard; ring-carrier adapters up to 300 mm available
- Chuck type & temp. → Gold-plated vacuum chuck, ambient → +130 °C with TC-2000 controller (standard “hot chuck”)
- Indexing accuracy/repeatability → ±4 µm X-Y, Z-resolution 0.25 mil (6.3 µm)
- Throughput → ≤ 1.5 s die-to-die at 5 mm step; lot-level throughput ≈ 60 wph (150 mm)
- Vision/alignment → PRM-3 CCD camera, Self-Teach Auto Alignment (STAA), optional OCR & Probe-Mark-Inspection (PMI)
- Interfaces → GPIB-IEEE 488, RS-232, Ethernet, EG-Commander/ProberBench software
Restrictions: Mechanical handling/alignment only (no electrical measurements); requires external tester Other: Modular design for cleanroom use; compatible with various probe cards
Documentation
Training required – contact lab staff (Chandan Ramakrishnaiah or Shivakumar Bhaskaran) Check lab resources for manuals (e.g., EG-Commander/ProberBench guides); no dedicated SOP listed – consider creating one
Recipes & Data
Standard Usage: Automated stepping for PCM/TEG lots; hot chuck for temp-dependent tests Process Control: Use consistent alignment routines; verify probe contact/scrub marks Notes: Interface with tester for full automation (GPIB/RS-232/Ethernet commands). Monitor chuck temp via TC-2000.